How do you calibrate the analog outputs on sc200 controller?
The shaded number indicates the digit that can be adjusted. Using the right or left arrow you can select a different digit. Using the up and down arrows, adjust the selected number on the SC200 display until the multimeter measures 20.00 mA. When the multimeter reads 20.00 mA, press the enter key on the SC200 to save this setting.
How to open a downloaded data file from the sc200 controller?
In the SD card directory, open the HACH folder. Select the Logs folder. Select a device folder. The event and data log files in the folder are shown. To view XML data log files: Make sure the HachDatalog.xsl style sheet exists in the device folder. Open the Excel application. Go to File, Open. Select the data log file.
How do I configure & verify the SC200 relay outputs? - Hach
The following are the basic steps for configuring the Alarm Relays on the SC200 controller: To configure the Alarm Relays: From the HOME screen, select MENU Select SC200 SETUP; Select RELAY SETUP; Select from: RELAY A; RELAY B; RELAY C; RELAY D; Select SOURCE; Select the sensor (by name or serial number) that this alarm should represent, and ...
How can you troubleshoot discrepancies between sensor ... - Hach
For sc200: 1. Output Cal 4 mA default is approximately +10500 ... Hach Small Business Supplier Program ...
How do I scale the 4-20 mA outputs on my sc200 controller to
The process for setting up and scaling the 4-20 mA outputs on the sc200 controller is as follows: Install probe on controller with "Scan Devices" within "Test/Maint" from the main menu. Proceed to "sc200 Setup" from the main menu. Choose "Output Setup" Choose the mA output that the SCADA is physically wired to (1 or 2)
How is the software for the sc200 updated? - Hach
2023年7月4日 · Software Update Procedure for sc200 Controller There are two ways to update the software for an sc200 controller. The one way is by using an SD card inserted directly into the controller and the second is using DataCom and a PC. First the latest version of the software for the sc200 Controller should be downloaded: sc200 Controller Software
Are there any instructions on saving data and event logs from an …
In the SD card directory, open the HACH folder. 3. Select the Logs folder. 4. Select a device folder. The event and data log files in the folder are shown. 5. To view XML data log files: a. Make sure the HachDatalog.xsl style sheet exists in the device folder. b. Open the Excel application. c. Go to File, Open. d. Select the data log file. e.
What SD cards can be used with the sc200? - Hach
SD cards can be used with the sc200. Answer. SD cards up to 32 GB can be used. ... ©2024 Hach All rights ...
How is the SC sensor or module software updated using the …
For the SC200 Controller, save the contents of the folder to the root directory of the SD card. Many of the sensors have an SD card image on the website. This image is specific to the SC200 file structure.NOTE: when using an SD card image it is best to save the file to a computer, unzip it and then transfer the entire HACH folder to the SD card.
What do the different icons on a SC200 controller mean? - Hach
Please note the below icons will be shown on both the main display of the SC200 and in the Diagnostics menu. Using the same icons in both locations can help a user understand what impact that diagnostic item has on the sensor and how to clear it.